Cliowin with HR-2000 ISA board error messages
Before proceeding further be sure to have Chapter 3 of CLIOwin User’s Manual in front of you.
CLIOwin with HR2000 ISA board error messages
Error
|
Possible cause
|
256
|
Physical I/O address doesn’t match Windows settings. This common error arises at the end of the installation when Windows has assigned an I/O address not matching the one set with jumper JP2. You can either move JP2 (see fig.3.1) to the other position or change (referring to fig.3.4) the Windows setting under: System Properties> Device Manager> Sound, video and game controller> Clio HR-2000> Properties> Resources> Input/Output Range> |
260
|
The hardware is not registered or disabled. Please execute the procedure described in 3.3. |
513
|
CLIO hardware not present or not working. This error is given when you remove the CLIO board from a computer where it was installed or when the board stops functioning. |
773
|
This error is given if you run CLIOwin in a computer where the hardware has not been installed. |